In this work we present the results of the experimental characterization of Silicon Drift Detectors (SDDs) readout by CUBE preamplifiers for X-ray spectroscopy measurements. One specific goal of the work was to characterize SDDs of different sizes cooled at low temperature in view of their use in the upgrade of the SIDDHARTA nuclear physics experiment. Beside the target application, the results of this work are also of interest for a more extended use of the SDDs in other X-ray spectroscopy applications. The SDDs have been designed as single units with square shape of different areas, 64 mm(2) (8 mm x 8 mm) and 144 mm(2) (12 mm x 12 mm ), and also as monolithic array 3 x 3 of elements of the 8 mm x 8 mm unit (total area 26 mm x 26 mm). The read-out of the SDDs is based on a CMOS (Complementary Metal Oxide Semiconductor) preamplifier (CUBE) both for the single unit and for the 3 x 3 array. For the readout of the array, an Application Specific Integrated Circuit (ASIC) has been used. An energy resolution better than 124 eV at the Mn - K alpha line has been measured with a 64 mm(2) SDD cooled at the temperature of 50 K. The energy resolution remains good (< 130 eV) also at short shaping time (250 ns) thanks to the noise feature of the CUBE preamplifier. Results of measurements on SDDs of different format and also on arrays of SDDs are presented in this work.

Silicon Drift Detectors and CUBE preamplifiers for high-resolution X-ray spectroscopy

QUAGLIA, RICCARDO;BOMBELLI, LUCA;BUSCA, PAOLO;FIORINI, CARLO ETTORE;OCCHIPINTI, MICHELE;
2015-01-01

Abstract

In this work we present the results of the experimental characterization of Silicon Drift Detectors (SDDs) readout by CUBE preamplifiers for X-ray spectroscopy measurements. One specific goal of the work was to characterize SDDs of different sizes cooled at low temperature in view of their use in the upgrade of the SIDDHARTA nuclear physics experiment. Beside the target application, the results of this work are also of interest for a more extended use of the SDDs in other X-ray spectroscopy applications. The SDDs have been designed as single units with square shape of different areas, 64 mm(2) (8 mm x 8 mm) and 144 mm(2) (12 mm x 12 mm ), and also as monolithic array 3 x 3 of elements of the 8 mm x 8 mm unit (total area 26 mm x 26 mm). The read-out of the SDDs is based on a CMOS (Complementary Metal Oxide Semiconductor) preamplifier (CUBE) both for the single unit and for the 3 x 3 array. For the readout of the array, an Application Specific Integrated Circuit (ASIC) has been used. An energy resolution better than 124 eV at the Mn - K alpha line has been measured with a 64 mm(2) SDD cooled at the temperature of 50 K. The energy resolution remains good (< 130 eV) also at short shaping time (250 ns) thanks to the noise feature of the CUBE preamplifier. Results of measurements on SDDs of different format and also on arrays of SDDs are presented in this work.
2015
Charge preamplifier; silicon drift detectors; X-ray spectroscopy; Electrical and Electronic Engineering; Nuclear Energy and Engineering; Nuclear and High Energy Physics; sezele
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/970168
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 42
  • ???jsp.display-item.citation.isi??? 39
social impact