In the timing of pulses, inhibition of the timing circuit during some time intervals (dead time effects, gating etc.), causes an error when the inhibition end falls during a pulse. This error is considered and a way to avoid it is discussed. Experimental results are shown and possible applications suggested.
On the precision timing of pulses, avoiding errors due to dead times or gate opening
COVA, SERGIO;BERTOLACCINI, MARIO
1970-01-01
Abstract
In the timing of pulses, inhibition of the timing circuit during some time intervals (dead time effects, gating etc.), causes an error when the inhibition end falls during a pulse. This error is considered and a way to avoid it is discussed. Experimental results are shown and possible applications suggested.File in questo prodotto: