Tricuspid regurgitation (TR) is the most common pathology of the tricuspid valve (TV), with significant mortality in severe cases. A well-established strategy to treat TR is represented by the clover surgical technique, which consists of stitching together the free edges of TV leaflets, producing a clover-shaped valvular orifice. Transcatheter treatments for TR constitute a valuable alternative for high-risk patients. In this work we investigated haemodynamic performances and safety of a novel device (StarTric device (STD)) aiming to perform the clover technique via percutaneous access. To assess haemodynamic performances, STD and clover were applied on porcine pathological TVs and tested. Fluid dynamic indexes of both strategies were compared to the pathological model. To evaluate device safety, forces exchanged between device and leaflets were compared to the extraction force (EF) required to STD to completely pass through the leaflet. Clover technique and STD induced a comparable TV backflow reduction (48% and 47%, respectively), with associated increase of TV flow in all tested conditions. Diastolic transvalvular pressure similarly increased indicating a reduction, though not significant, of the valvular orifice. Forces ranged from 1N to 1.71N, compared to an EF of 22.16 +/- 8.6N. Force varied significantly amongst different working conditions (normotensive, mild, and severe hypertensive) for each leaflet, whilst no significative variation was found on different leaflets in the same working condition. In the adopted experimental scenario, STD demonstrated comparable efficacy to the surgical strategy in restoring TV haemodynamic. The forces acting on the leaflets following STD implantation were far lower when compared to EFs.
A Novel Transcatheter Device for the Edge-to-Edge Treatment of Tricuspid Regurgitation: A Preliminary Evaluation
Salurso, Eleonora;Perico, Francesca;Pappalardo, Fabio;Vismara, Riccardo
2023-01-01
Abstract
Tricuspid regurgitation (TR) is the most common pathology of the tricuspid valve (TV), with significant mortality in severe cases. A well-established strategy to treat TR is represented by the clover surgical technique, which consists of stitching together the free edges of TV leaflets, producing a clover-shaped valvular orifice. Transcatheter treatments for TR constitute a valuable alternative for high-risk patients. In this work we investigated haemodynamic performances and safety of a novel device (StarTric device (STD)) aiming to perform the clover technique via percutaneous access. To assess haemodynamic performances, STD and clover were applied on porcine pathological TVs and tested. Fluid dynamic indexes of both strategies were compared to the pathological model. To evaluate device safety, forces exchanged between device and leaflets were compared to the extraction force (EF) required to STD to completely pass through the leaflet. Clover technique and STD induced a comparable TV backflow reduction (48% and 47%, respectively), with associated increase of TV flow in all tested conditions. Diastolic transvalvular pressure similarly increased indicating a reduction, though not significant, of the valvular orifice. Forces ranged from 1N to 1.71N, compared to an EF of 22.16 +/- 8.6N. Force varied significantly amongst different working conditions (normotensive, mild, and severe hypertensive) for each leaflet, whilst no significative variation was found on different leaflets in the same working condition. In the adopted experimental scenario, STD demonstrated comparable efficacy to the surgical strategy in restoring TV haemodynamic. The forces acting on the leaflets following STD implantation were far lower when compared to EFs.File | Dimensione | Formato | |
---|---|---|---|
11311-1258848_Salurso.pdf
accesso aperto
:
Publisher’s version
Dimensione
1.61 MB
Formato
Adobe PDF
|
1.61 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.