To improve non-invasively the specificity in the diagnosis of breast cancer after a positive screening mammography or doubt/suspicious ultrasound examination, the SOLUS project developed a multimodal imaging system that combines: B-mode ultrasound (US) scans (to assess morphology), Color Doppler (to visualize vascularization), shear-wave elastography (to measure stiffness), and time domain multi-wavelength diffuse optical tomography (to estimate tissue composition in terms of oxy- and deoxy-hemoglobin, lipid, water, and collagen concentrations). The multimodal probe arranges 8 innovative photonic modules (optodes) around the US transducer, providing capability for optical tomographic reconstruction. For more accurate estimate of lesion composition, US-assessed morphological priors can be used to guide the optical reconstructions. Each optode comprises: i) 8 picosecond pulsed laser diodes with different wavelengths, covering a wide spectral range (635-1064 nm) for good probing of the different tissue constituents; ii) a large-area (variable, up to 8.6 mm2) fast-gated digital Silicon Photomultiplier; iii) the acquisition electronics to record the distribution of time-of-flight of the re-emitted photons. The optode is the basic element of the optical part of the system, but is also a stand-alone, ultra-compact (about 4 cm3) device for time domain multi-wavelength diffuse optics, with potential application in various fields.

SOLUS: An innovative multimodal imaging system to improve breast cancer diagnosis through diffuse optics and ultrasounds

Pifferi A.;Dalla Mora A.;Di Sieno L.;Ferocino E.;Tosi A.;Conca E.;Sesta V.;Taroni P.
2021-01-01

Abstract

To improve non-invasively the specificity in the diagnosis of breast cancer after a positive screening mammography or doubt/suspicious ultrasound examination, the SOLUS project developed a multimodal imaging system that combines: B-mode ultrasound (US) scans (to assess morphology), Color Doppler (to visualize vascularization), shear-wave elastography (to measure stiffness), and time domain multi-wavelength diffuse optical tomography (to estimate tissue composition in terms of oxy- and deoxy-hemoglobin, lipid, water, and collagen concentrations). The multimodal probe arranges 8 innovative photonic modules (optodes) around the US transducer, providing capability for optical tomographic reconstruction. For more accurate estimate of lesion composition, US-assessed morphological priors can be used to guide the optical reconstructions. Each optode comprises: i) 8 picosecond pulsed laser diodes with different wavelengths, covering a wide spectral range (635-1064 nm) for good probing of the different tissue constituents; ii) a large-area (variable, up to 8.6 mm2) fast-gated digital Silicon Photomultiplier; iii) the acquisition electronics to record the distribution of time-of-flight of the re-emitted photons. The optode is the basic element of the optical part of the system, but is also a stand-alone, ultra-compact (about 4 cm3) device for time domain multi-wavelength diffuse optics, with potential application in various fields.
2021
Progress in Biomedical Optics and Imaging - Proceedings of SPIE
9781510641136
9781510641143
Breast cancer
Diffuse optical tomography
Diffuse optics
Multimodal imaging
Optical mammography
Shear wave elastography
Ultrasound
sezele
File in questo prodotto:
File Dimensione Formato  
116390C.pdf

Accesso riservato

Descrizione: Full text
: Publisher’s version
Dimensione 397.26 kB
Formato Adobe PDF
397.26 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1186232
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? ND
social impact