It has been brought to our attention that there is ambiguity in the ERS technical standards for respiratory oscillometry, specifically regarding the use of test loads for calibration and verification. Here it is worth making a distinction between recommendations for manufacturers versus users. We recognise that a standardised reactive test load has not been established for calibration or verification. This is an active research area, driven by the poor comparability in reactance which remains across devices.
Technical standards for respiratory oscillometry: test loads for calibration and verification
Dellacà, Raffaele;
2020-01-01
Abstract
It has been brought to our attention that there is ambiguity in the ERS technical standards for respiratory oscillometry, specifically regarding the use of test loads for calibration and verification. Here it is worth making a distinction between recommendations for manufacturers versus users. We recognise that a standardised reactive test load has not been established for calibration or verification. This is an active research area, driven by the poor comparability in reactance which remains across devices.File in questo prodotto: