This study deals with a microbial air sampler which differently from the most used ones, is using a centrifugal effect within a spiral coil. Physical collection efficiency and cut-off size (d50) are investigated through a DPM CFD model, a simplified analytical model and some experimental tests. The results are compared and discussed at different airflow rates. A quite good agreement is found for diameter ≥ 2 μm.
Physical collection efficiency of a single stage centrifugal air sampler. Analytical, CFD and experimental results
ROMANO, FRANCESCO;MAROCCO, LUCA DAVIDE;JOPPOLO, CESARE MARIA
2016-01-01
Abstract
This study deals with a microbial air sampler which differently from the most used ones, is using a centrifugal effect within a spiral coil. Physical collection efficiency and cut-off size (d50) are investigated through a DPM CFD model, a simplified analytical model and some experimental tests. The results are compared and discussed at different airflow rates. A quite good agreement is found for diameter ≥ 2 μm.File in questo prodotto:
File | Dimensione | Formato | |
---|---|---|---|
Physical collection efficiency of a single stage centrifugal air sampler. Analytical, CFD and experimental results.pdf
Accesso riservato
Descrizione: Articolo principale
:
Publisher’s version
Dimensione
793.74 kB
Formato
Adobe PDF
|
793.74 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.