A theoretical and experimental comparison between vertical and coplanar interdigitated sensing configurations for impedimetric cell growth tracking is presented. These widely-adopted approaches are quantitatively compared on the same cell populations and on the same 10 μm interdigitated microelectrodes using a versatile custom-made monitoring platform including a 24-channel miniaturized potentiostat. The characterization of bare microelectrodes in buffer and tracking experiments with HeLa cells over 16 hours demonstrate that the coplanar configuration provides a higher sensitivity to cell adhesion and spreading (Cell Index = 1.6 vs. 0.4) albeit at a higher frequency of maximum sensitivity (100 kHz vs. 24 kHz) shifting over time.
Theoretical and experimental comparison of microelectrode sensing configurations for impedimetric cell monitoring
CARMINATI, MARCO;FERRARI, GIORGIO;SAMPIETRO, MARCO;
2013-01-01
Abstract
A theoretical and experimental comparison between vertical and coplanar interdigitated sensing configurations for impedimetric cell growth tracking is presented. These widely-adopted approaches are quantitatively compared on the same cell populations and on the same 10 μm interdigitated microelectrodes using a versatile custom-made monitoring platform including a 24-channel miniaturized potentiostat. The characterization of bare microelectrodes in buffer and tracking experiments with HeLa cells over 16 hours demonstrate that the coplanar configuration provides a higher sensitivity to cell adhesion and spreading (Cell Index = 1.6 vs. 0.4) albeit at a higher frequency of maximum sensitivity (100 kHz vs. 24 kHz) shifting over time.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.