Plantar pressure measurements provide useful information to diagnose a diverse range of foot disorders; unfortunately, the commercially available measurement systems are undesirably sensitive to several disturbances, but this aspect is mostly neglected in the literature. This paper describes the results of an experimental campaign aiming at the identification of pressure measuring system metrological performances, at system modeling, and at the implementation of correction procedures. The sensor model was implemented using the results of static and dynamic tests performed on a pedar-X plantar pressure measurement system. The static calibration was performed by analyzing the effect of temperature, single sensor coverage area, local curvature, tangential forces, long-term stability (creep), and sensor crosstalk on the system performances. The dynamic calibration was performed on an electrodynamic shaker, identifying the single sensor frequency response function and the hysteresis under different average loads. The dynamic sensor model is based on the Kelvin-Voigt model, which is representative of the viscoelastic behavior of the material. The model allowed us to compensate both the creep (i.e., the behavior under static loads) and the nonunitary frequency response function. A deconvolution-based algorithm has been proposed to compensate the sensor crosstalk effects, although its implementation requires additional investigations. Experimental results of bobbing and gait tests showed that, with the adoption of the proposed compensation algorithms, the force and center of pressure errors could be reduced by more than 50\% of their initial values.

Metrological performances of a plantar pressure measurement system

SAGGIN, BORTOLINO;SCACCABAROZZI, DIEGO;TARABINI, MARCO
2013-01-01

Abstract

Plantar pressure measurements provide useful information to diagnose a diverse range of foot disorders; unfortunately, the commercially available measurement systems are undesirably sensitive to several disturbances, but this aspect is mostly neglected in the literature. This paper describes the results of an experimental campaign aiming at the identification of pressure measuring system metrological performances, at system modeling, and at the implementation of correction procedures. The sensor model was implemented using the results of static and dynamic tests performed on a pedar-X plantar pressure measurement system. The static calibration was performed by analyzing the effect of temperature, single sensor coverage area, local curvature, tangential forces, long-term stability (creep), and sensor crosstalk on the system performances. The dynamic calibration was performed on an electrodynamic shaker, identifying the single sensor frequency response function and the hysteresis under different average loads. The dynamic sensor model is based on the Kelvin-Voigt model, which is representative of the viscoelastic behavior of the material. The model allowed us to compensate both the creep (i.e., the behavior under static loads) and the nonunitary frequency response function. A deconvolution-based algorithm has been proposed to compensate the sensor crosstalk effects, although its implementation requires additional investigations. Experimental results of bobbing and gait tests showed that, with the adoption of the proposed compensation algorithms, the force and center of pressure errors could be reduced by more than 50\% of their initial values.
File in questo prodotto:
File Dimensione Formato  
ieeeMetrologicalPressureMaps.pdf

Accesso riservato

: Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione 1.24 MB
Formato Adobe PDF
1.24 MB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/719955
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 38
  • ???jsp.display-item.citation.isi??? 29
social impact