In-phantom absorbed dose is imaged by measuring the optical transmittance through ferrous-sulphate and xylenolorange infused gel, in form of layers. Images are acquired with a CCD camera provided with a suitable optical filter. The dependence of dosimeter sensitivity and of linearity range on the amount of xylenol orange and on parameters of optical analysis has been inspected. In order to check the capability of the method, dose profiles in the central axis of a phantom irradiated with a collimated c-beam have been compared with the results obtained with a ionisation chamber.
Study of Light Transmittance from Layers of Fricke-xylenol-orange-gel Dosimeters
MARIANI, MARIO;
2004-01-01
Abstract
In-phantom absorbed dose is imaged by measuring the optical transmittance through ferrous-sulphate and xylenolorange infused gel, in form of layers. Images are acquired with a CCD camera provided with a suitable optical filter. The dependence of dosimeter sensitivity and of linearity range on the amount of xylenol orange and on parameters of optical analysis has been inspected. In order to check the capability of the method, dose profiles in the central axis of a phantom irradiated with a collimated c-beam have been compared with the results obtained with a ionisation chamber.File in questo prodotto:
File | Dimensione | Formato | |
---|---|---|---|
MARIANI Study Light Trans Gel NIM04.pdf
Accesso riservato
:
Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione
318.37 kB
Formato
Adobe PDF
|
318.37 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.