Thrombus aspiration catheters are devices used to remove a blood clot from a vessel, usually prior to angioplasty or stent implantation. However, in vitro results showed that the use of different commercial devices could produce very different thrombus removals, suggesting a primary dependence on the distal tip configuration of the catheter.A computational methodology based on realistic catheter tip modeling was developed to investigate the factors affecting the thrombus suction. Two different designs were considered, either with a single central lumen or a combination of central and side holes. First, steadystate aspiration of distilled water from a reservoir was simulated and compared with experimental tests. Subsequently, the aspiration of a totally occlusive thrombus, modeled as a high viscous fluid, was simulated solving a complex two-phase (blood and thrombus) problem. In particular, the benefit of additional openings was investigated. Good matching between the steady-state experimental and numerically simulated hydraulic behaviors allowed a validation of the numerical models. Numerical results of thrombus aspiration showed that the catheter with central and side holes had a worse performance if compared with the single central lumen catheter. Indeed, the inlets in contact with both blood and thrombus preferentially aspirate blood due to its much lower viscosity. This effect hindered the aspiration of thrombus.The amount of aspirated thrombus highly depends on the complex, two-phase fluid dynamics occurring across the catheter tips. Results suggested that location of additional holes is crucial in the catheter aspiration performance.

Numerical Simulation of Thrombus Aspiration in TwoRealistic Models of Catheter Tips

PENNATI, GIANCARLO;DUBINI, GABRIELE ANGELO;MIGLIAVACCA, FRANCESCO
2010-01-01

Abstract

Thrombus aspiration catheters are devices used to remove a blood clot from a vessel, usually prior to angioplasty or stent implantation. However, in vitro results showed that the use of different commercial devices could produce very different thrombus removals, suggesting a primary dependence on the distal tip configuration of the catheter.A computational methodology based on realistic catheter tip modeling was developed to investigate the factors affecting the thrombus suction. Two different designs were considered, either with a single central lumen or a combination of central and side holes. First, steadystate aspiration of distilled water from a reservoir was simulated and compared with experimental tests. Subsequently, the aspiration of a totally occlusive thrombus, modeled as a high viscous fluid, was simulated solving a complex two-phase (blood and thrombus) problem. In particular, the benefit of additional openings was investigated. Good matching between the steady-state experimental and numerically simulated hydraulic behaviors allowed a validation of the numerical models. Numerical results of thrombus aspiration showed that the catheter with central and side holes had a worse performance if compared with the single central lumen catheter. Indeed, the inlets in contact with both blood and thrombus preferentially aspirate blood due to its much lower viscosity. This effect hindered the aspiration of thrombus.The amount of aspirated thrombus highly depends on the complex, two-phase fluid dynamics occurring across the catheter tips. Results suggested that location of additional holes is crucial in the catheter aspiration performance.
2010
Thrombus aspiration; Catheter performance; Cardiology intervention; Computational fluid dynamics; Side holes; Two-phase flow
File in questo prodotto:
File Dimensione Formato  
Pennati 2010 Artificail Organs.pdf

Accesso riservato

: Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione 195.47 kB
Formato Adobe PDF
195.47 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/569689
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 21
  • ???jsp.display-item.citation.isi??? 19
social impact