Aim of this study is to develop a mechanical simulator (MS) reproducing cardiac wall kinematics [i.e., radial (R), longitudinal (L) and rotational (RT) motions] to test piezoelectric gyroscopic sensors (GS) that are able to measure cardiac torsion that has proved to be a sensitive index of cardiac performance. The MS consists of three brushless motors controlled by a dedicated software either separately or simultaneously reproducing the three main cardiac wall movements (R, L, RT) obtained by implementing different physiologic or pathologic velocity profiles derived from in vivo data. GS accuracy (max % error) was experimentally tested by connecting it to the MS driven in velocity in different working conditions [i.e., cardiac period (515-1030 ms), RT angle (4-16 degrees), GS axis inclination (0-90 degrees) with respect to the cardiac rotation axis]. The MS reproduced the tested velocity profiles well. The GS showed high accuracy in measuring both physiologic and pathologic RT velocity profiles, whereas they proved insensitive to R and L motions. GS axis inclination influenced measurements; however, it was possible to correct this taking the inclination angle cosine into account. The MS proved to be a useful tool to study cardiac wall kinematics and test GS reliability with a view to in vivo application.

A mechanical simulator of cardiac wall kinematics

CUTRI', ELENA;BAGNOLI, PAOLA;COSTANTINO, MARIA LAURA;FUMERO, ROBERTO
2010-01-01

Abstract

Aim of this study is to develop a mechanical simulator (MS) reproducing cardiac wall kinematics [i.e., radial (R), longitudinal (L) and rotational (RT) motions] to test piezoelectric gyroscopic sensors (GS) that are able to measure cardiac torsion that has proved to be a sensitive index of cardiac performance. The MS consists of three brushless motors controlled by a dedicated software either separately or simultaneously reproducing the three main cardiac wall movements (R, L, RT) obtained by implementing different physiologic or pathologic velocity profiles derived from in vivo data. GS accuracy (max % error) was experimentally tested by connecting it to the MS driven in velocity in different working conditions [i.e., cardiac period (515-1030 ms), RT angle (4-16 degrees), GS axis inclination (0-90 degrees) with respect to the cardiac rotation axis]. The MS reproduced the tested velocity profiles well. The GS showed high accuracy in measuring both physiologic and pathologic RT velocity profiles, whereas they proved insensitive to R and L motions. GS axis inclination influenced measurements; however, it was possible to correct this taking the inclination angle cosine into account. The MS proved to be a useful tool to study cardiac wall kinematics and test GS reliability with a view to in vivo application.
2010
cardiac kinematics; mechanical simulator; cariac torsion
File in questo prodotto:
File Dimensione Formato  
2010-3-ASAIO.pdf

Accesso riservato

: Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione 1.09 MB
Formato Adobe PDF
1.09 MB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/561825
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 6
social impact