Bonner sphere spectrometers (BSS) are being used widely in neutron spectrometry since 1960. The response to neutrons of these moderating detectors spans over a broad energy range. This work discusses the measurement of the sensitivity of an extended range BSS hosting a CR39 nuclear track detector coupled to a boron converter. Two kinds of boron converters were tested, the first made by natural boron and a second one enriched in 10B. The BSS response functions were calculated with Monte Carlo simulation using the MCNPX 2.4.0. and the FLUKA codes.
Sensitivity Study of CR39 Track Detector in an Extended Range Bonner Sphere Spectrometer
CARESANA, MARCO;GIULINI CASTIGLIONI AGOSTEO, STEFANO LUIGI MARIA;CAMPI, FABRIZIO;FERRARINI, MICHELE;PORTA, ALESSANDRO ANTONIO;
2007-01-01
Abstract
Bonner sphere spectrometers (BSS) are being used widely in neutron spectrometry since 1960. The response to neutrons of these moderating detectors spans over a broad energy range. This work discusses the measurement of the sensitivity of an extended range BSS hosting a CR39 nuclear track detector coupled to a boron converter. Two kinds of boron converters were tested, the first made by natural boron and a second one enriched in 10B. The BSS response functions were calculated with Monte Carlo simulation using the MCNPX 2.4.0. and the FLUKA codes.File in questo prodotto:
File | Dimensione | Formato | |
---|---|---|---|
2007 sfere di bonner rad prot dosimetry.pdf
Accesso riservato
:
Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione
155.33 kB
Formato
Adobe PDF
|
155.33 kB | Adobe PDF | Visualizza/Apri |