Background and purpose - Safety notices for medical devices such as total knee arthroplasty (TKA) implants indicate problems in their design or performance that require corrective action to prevent patient harm. Safety notices often published on national Ministries of Health or regulatory agencies websites. It is unknown whether problems triggering safety notices identify the same implants as those identified by registries as "outlier." We aimed to assess extent to which safety notices and outlier identification registries signal the same or different TKA implants. Methods - The CORE-MD tool, an automated scraper tool, was used to collect safety notices related TKA implants on 13 national Ministries of Health websites and regulatory agencies. Safety notices were defined according to the Medical Device Regulation (MDR) as "a communication sent by a manufacturer to users or customers relation to a field safety corrective action." Identified TKA outliers, defined as having a significantly higher revision than other comparable TKA implants, were extracted from registry reports. Results - 787 safety notices for 38 TKA implants 35 TKA outliers were identified, together identifying unique TKA implants. 26 (55%) TKA implants had safety notices and were also outliers, 12 (26%) TKA implants only safety notices, and 9 (19%) were outliers only. TKA implants with safety notices only had similar types of problems to TKA outliers with safety notices, with "Manufacturing/Packaging/Shipping" problems being most frequent (44%). Cumulative revision risks (1/5/10 years) were lower for TKA implants with safety notices only than for TKA outliers with safety notices. Conclusion - 55% of the TKA with a safety notice were identified as outliers in the registry, whereas around 25% of TKA outliers are not the subject of publicly released safety notices, with safety notices pointing to TKA implants not identified by registries as potentially having a higher risk of failure. This suggests that safety notices and registry outlier data measure different aspects of safety and performance.

Safety notices and registry outlier data measure different aspects of safety and performance of total knee implants: a comparative study of safety notices and register outliers

Ren, Yijun;Caiani, Enrico G;
2024-01-01

Abstract

Background and purpose - Safety notices for medical devices such as total knee arthroplasty (TKA) implants indicate problems in their design or performance that require corrective action to prevent patient harm. Safety notices often published on national Ministries of Health or regulatory agencies websites. It is unknown whether problems triggering safety notices identify the same implants as those identified by registries as "outlier." We aimed to assess extent to which safety notices and outlier identification registries signal the same or different TKA implants. Methods - The CORE-MD tool, an automated scraper tool, was used to collect safety notices related TKA implants on 13 national Ministries of Health websites and regulatory agencies. Safety notices were defined according to the Medical Device Regulation (MDR) as "a communication sent by a manufacturer to users or customers relation to a field safety corrective action." Identified TKA outliers, defined as having a significantly higher revision than other comparable TKA implants, were extracted from registry reports. Results - 787 safety notices for 38 TKA implants 35 TKA outliers were identified, together identifying unique TKA implants. 26 (55%) TKA implants had safety notices and were also outliers, 12 (26%) TKA implants only safety notices, and 9 (19%) were outliers only. TKA implants with safety notices only had similar types of problems to TKA outliers with safety notices, with "Manufacturing/Packaging/Shipping" problems being most frequent (44%). Cumulative revision risks (1/5/10 years) were lower for TKA implants with safety notices only than for TKA outliers with safety notices. Conclusion - 55% of the TKA with a safety notice were identified as outliers in the registry, whereas around 25% of TKA outliers are not the subject of publicly released safety notices, with safety notices pointing to TKA implants not identified by registries as potentially having a higher risk of failure. This suggests that safety notices and registry outlier data measure different aspects of safety and performance.
2024
File in questo prodotto:
File Dimensione Formato  
17858-HOOGERVORST+D.pdf

accesso aperto

Descrizione: paper
: Publisher’s version
Dimensione 578.62 kB
Formato Adobe PDF
578.62 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1279939
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact