A robotic cone-beam computed tomography (CT) scanner has inherent advantages. In particular, it enables source and detector trajectories capable of extending the field-of-view (FOV) of the reconstructed CT images, where the FOV is defined as the region in the source trajectory plane for which all ray lines are acquired (at any position and direction in this plane). Previous extensions of the FOV used an offset detector or a displaced center of rotation and a single full 360◦ scan. However, due to limitations in the conventional range of motion inside the treatment room to avoid collisions, some systems can only perform short scans. This paper investigates a new approach to FOV extension for an existing system by adapting the FOV extension to two complementary short scans with displaced centers of rotation. We validate this approach on numerical simulations of the Forbild thorax phantom.

Extension of the cone-beam CT field-of-view using two short scans with displaced centers of rotation

Belotti G.;Baroni G.
2022-01-01

Abstract

A robotic cone-beam computed tomography (CT) scanner has inherent advantages. In particular, it enables source and detector trajectories capable of extending the field-of-view (FOV) of the reconstructed CT images, where the FOV is defined as the region in the source trajectory plane for which all ray lines are acquired (at any position and direction in this plane). Previous extensions of the FOV used an offset detector or a displaced center of rotation and a single full 360◦ scan. However, due to limitations in the conventional range of motion inside the treatment room to avoid collisions, some systems can only perform short scans. This paper investigates a new approach to FOV extension for an existing system by adapting the FOV extension to two complementary short scans with displaced centers of rotation. We validate this approach on numerical simulations of the Forbild thorax phantom.
2022
Proc. SPIE 12304, 7th International Conference on Image Formation in X-Ray Computed Tomography, 123040E
9781510656697
9781510656703
CBCT, x-ray, partial scan, FDK, dual scan
File in questo prodotto:
File Dimensione Formato  
123040E.pdf

accesso aperto

Descrizione: Articolo
: Publisher’s version
Dimensione 3.1 MB
Formato Adobe PDF
3.1 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1231414
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? ND
social impact