This paper summarizes the design implementation and validation of the measurement setup for the thermal characterization of the thermally insulating mountings of the MAJIS instrument. The challenges of the measurement derive from the low value of the conductance to be measured, in the range of few mW/K, the wide temperature range to be investigated, from 100 to 300 K, the need of operating under vacuum. The developed configuration exploits the guarded plate configuration to reduce the uncertainty on the parasitic heat fluxes, multilayer insulators to minimize the radiative heat exchange, a liquid nitrogen loop for the cooling of the heat sink. The setup has been built and validated with tests on a specimen to assess the repeatability of the measurement. With the characterization of different specimens from the same production batch the setup has allowed also assessing the repeatability of the manufacturing process.

Thermal conductance measurement of the MAJIS insulating mountings down to cryogenic temperature

Saggin B.;Scaccabarozzi D.;Moschioni G.;
2022-01-01

Abstract

This paper summarizes the design implementation and validation of the measurement setup for the thermal characterization of the thermally insulating mountings of the MAJIS instrument. The challenges of the measurement derive from the low value of the conductance to be measured, in the range of few mW/K, the wide temperature range to be investigated, from 100 to 300 K, the need of operating under vacuum. The developed configuration exploits the guarded plate configuration to reduce the uncertainty on the parasitic heat fluxes, multilayer insulators to minimize the radiative heat exchange, a liquid nitrogen loop for the cooling of the heat sink. The setup has been built and validated with tests on a specimen to assess the repeatability of the measurement. With the characterization of different specimens from the same production batch the setup has allowed also assessing the repeatability of the manufacturing process.
2022
2022 IEEE 9th International Workshop on Metrology for AeroSpace, MetroAeroSpace 2022 - Proceedings
978-1-6654-1076-2
accuracy
cryogenic insulators
guarded plate
Thermal conductance
thermal resistance
uncertainty
File in questo prodotto:
File Dimensione Formato  
Paper.pdf

Accesso riservato

: Altro materiale allegato
Dimensione 396.07 kB
Formato Adobe PDF
396.07 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1221490
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact