The purpose of this work is to propose a workflow that couples experimental and computational activities aimed at developing a credible digital twin of a commercial coronary bioresorbable vascular scaffold when direct access to data about material mechanical properties is not possible. Such a situation is be faced when the manufacturer is not involved in the study, thus directly investigating the actual device is the only source of information available. The object of the work is the Fantom® Encore polymeric stent (REVA Medical) made of TyrocoreTM. Four devices were purchased and used in mechanical tests that are easily reproducible in any mechanical laboratory, i.e. free expansion and uniaxial tension testing, the latter performed with protocols that emphasized the rate-dependent properties of the polymer. Given the complexity of the mechanical behaviour observed experimentally, it was chosen to use the Parallel Rehological Framework material model, already used in the literature to describe the behaviour of other polymers, such as PLLA. Calibration of the material model was based on simulations that replicate the tensile test performed on the device. Given the high number of material parameters, a plan of simulations was done to find the most suitable set, varying each parameter value in a feasible range and considering a single repetitive unit of the stent, neglecting residual stresses generated by crimping and expansion. This strategy resulted in a significant reduction of computational cost. The performance of the set of parameters thus identified was finally evaluated considering the whole delivery system, by comparing the experimental results with the data collected simulating free expansion and uniaxial tension testing. Moreover, radial force testing was numerically performed and compared with literature data. The obtained results demonstrated the effectiveness of the digital twin development pipeline, a path applicable to any commercial device whose geometric structure is based on repetitive units.

From the real device to the digital twin: A coupled experimental-numerical strategy to investigate a novel bioresorbable vascular scaffold

Antonini L.;Berti F.;Isella B.;Hossain D.;Mandelli L.;Pennati G.;Petrini L.
2021-01-01

Abstract

The purpose of this work is to propose a workflow that couples experimental and computational activities aimed at developing a credible digital twin of a commercial coronary bioresorbable vascular scaffold when direct access to data about material mechanical properties is not possible. Such a situation is be faced when the manufacturer is not involved in the study, thus directly investigating the actual device is the only source of information available. The object of the work is the Fantom® Encore polymeric stent (REVA Medical) made of TyrocoreTM. Four devices were purchased and used in mechanical tests that are easily reproducible in any mechanical laboratory, i.e. free expansion and uniaxial tension testing, the latter performed with protocols that emphasized the rate-dependent properties of the polymer. Given the complexity of the mechanical behaviour observed experimentally, it was chosen to use the Parallel Rehological Framework material model, already used in the literature to describe the behaviour of other polymers, such as PLLA. Calibration of the material model was based on simulations that replicate the tensile test performed on the device. Given the high number of material parameters, a plan of simulations was done to find the most suitable set, varying each parameter value in a feasible range and considering a single repetitive unit of the stent, neglecting residual stresses generated by crimping and expansion. This strategy resulted in a significant reduction of computational cost. The performance of the set of parameters thus identified was finally evaluated considering the whole delivery system, by comparing the experimental results with the data collected simulating free expansion and uniaxial tension testing. Moreover, radial force testing was numerically performed and compared with literature data. The obtained results demonstrated the effectiveness of the digital twin development pipeline, a path applicable to any commercial device whose geometric structure is based on repetitive units.
2021
Angioplasty, Balloon, Coronary
Coronary Vessels
Drug-Eluting Stents
Prosthesis Design
Absorbable Implants
File in questo prodotto:
File Dimensione Formato  
A136 InSilc - Antonini REVA - PLOSone 2021.pdf

accesso aperto

Descrizione: Articolo
: Publisher’s version
Dimensione 2.53 MB
Formato Adobe PDF
2.53 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1205194
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 7
social impact