As key cellular elements of hemostasis, platelets represent a primary target for thrombosis and bleeding management. Currently, therapeutic manipulations of platelet function (antithrombotic drugs) and count (platelet transfusion) are performed with limited or no real-time monitoring of the desired outcome at the point-of-care. To address the need, we have designed and fabricated an easy-to-use, accurate, and portable impedance aggregometer called “MICELI” (MICrofluidic, ELectrical, Impedance). It improves on current platelet aggregation technology by decreasing footprint, assay complexity, and time to obtain results. The current study aimed to optimize the MICELI protocol; validate sensitivity to aggregation agonists and key blood parameters, i.e., platelet count and hematocrit; and verify the MICELI operational performance as compared to commercial impedance aggregometry. We demonstrated that the MICELI aggregometer could detect platelet aggregation in 250 µL of whole blood or platelet-rich plasma, stimulated by ADP, TRAP-6, collagen, epinephrine, and calcium ionophore. Using hirudin as blood anticoagulant allowed higher aggregation values. Aggregation values obtained by the MICELI strongly correlated with platelet count and were not affected by hematocrit. The operational performance comparison of the MICELI and the Multiplate® Analyzer demonstrated strong correlation and similar interdonor distribution of aggregation values obtained between these devices. With the proven reliability of the data obtained by the MICELI aggregometer, it can be further translated into a point-of-care diagnostic device aimed at monitoring platelet function in order to guide pharmacological hemostasis management and platelet transfusions.
The MICELI (MICrofluidic, ELectrical, impedance): Prototyping a point-of-care impedance platelet aggregometer
Bozzi S.;Dimasi A.;Rasponi M.;Santoleri A.;Consolo F.;Redaelli A.
2020-01-01
Abstract
As key cellular elements of hemostasis, platelets represent a primary target for thrombosis and bleeding management. Currently, therapeutic manipulations of platelet function (antithrombotic drugs) and count (platelet transfusion) are performed with limited or no real-time monitoring of the desired outcome at the point-of-care. To address the need, we have designed and fabricated an easy-to-use, accurate, and portable impedance aggregometer called “MICELI” (MICrofluidic, ELectrical, Impedance). It improves on current platelet aggregation technology by decreasing footprint, assay complexity, and time to obtain results. The current study aimed to optimize the MICELI protocol; validate sensitivity to aggregation agonists and key blood parameters, i.e., platelet count and hematocrit; and verify the MICELI operational performance as compared to commercial impedance aggregometry. We demonstrated that the MICELI aggregometer could detect platelet aggregation in 250 µL of whole blood or platelet-rich plasma, stimulated by ADP, TRAP-6, collagen, epinephrine, and calcium ionophore. Using hirudin as blood anticoagulant allowed higher aggregation values. Aggregation values obtained by the MICELI strongly correlated with platelet count and were not affected by hematocrit. The operational performance comparison of the MICELI and the Multiplate® Analyzer demonstrated strong correlation and similar interdonor distribution of aggregation values obtained between these devices. With the proven reliability of the data obtained by the MICELI aggregometer, it can be further translated into a point-of-care diagnostic device aimed at monitoring platelet function in order to guide pharmacological hemostasis management and platelet transfusions.File | Dimensione | Formato | |
---|---|---|---|
ijms-21-01174-v2.pdf
accesso aperto
:
Publisher’s version
Dimensione
4.92 MB
Formato
Adobe PDF
|
4.92 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.