Ballistocardiography (BCG), measuring the heart beat-induced mass movements of the body generated by forces associated to heart contraction, has been recently reconsidered as a useful non-invasive technique to characterize cardiac performance, potentially offering a simple, efficient and affordable solution for e-health monitoring. Our aim was to extract heart rate variability indices from BCG signal (B-HRV), and to evaluate their correlation with ECG-derived (E-HRV) ones, both in time and frequency domains. Five-minute BCG signals from 18 healthy volunteers were acquired by a piezoelectric 3D-force plate simultaneously with the ECG signal, in both standing (ST) and supine (SUP) positions. BCG systolic waves (H, I, J and K) in the longitudinal (head-foot) component were detected, from which B-HRV series were computed and compared by correlation and Bland-Altman analysis against conventional RR. Posture affected accuracy of B-HRV time and frequency parameters compared to ECG-HRV, with the best performance obtained in SUP by the JJ series. This signal could be used as an alternative measurement of HRV in novel home e-health monitoring systems.

Studying heart rate variability from ballistocardiography acquired by force platform: Comparison with conventional ECG

Martín-Yebra, Alba;Landreani, Federica;Casellato, Claudia;Pavan, Esteban;Frigo, Carlo;Caiani, Enrico G.
2015-01-01

Abstract

Ballistocardiography (BCG), measuring the heart beat-induced mass movements of the body generated by forces associated to heart contraction, has been recently reconsidered as a useful non-invasive technique to characterize cardiac performance, potentially offering a simple, efficient and affordable solution for e-health monitoring. Our aim was to extract heart rate variability indices from BCG signal (B-HRV), and to evaluate their correlation with ECG-derived (E-HRV) ones, both in time and frequency domains. Five-minute BCG signals from 18 healthy volunteers were acquired by a piezoelectric 3D-force plate simultaneously with the ECG signal, in both standing (ST) and supine (SUP) positions. BCG systolic waves (H, I, J and K) in the longitudinal (head-foot) component were detected, from which B-HRV series were computed and compared by correlation and Bland-Altman analysis against conventional RR. Posture affected accuracy of B-HRV time and frequency parameters compared to ECG-HRV, with the best performance obtained in SUP by the JJ series. This signal could be used as an alternative measurement of HRV in novel home e-health monitoring systems.
2015
2015 COMPUTING IN CARDIOLOGY CONFERENCE (CINC)
9781509006854
Computer Science (all); Cardiology and Cardiovascular Medicine
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1084354
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 15
social impact