The Italian National Centre for Oncological Hadrontherapy (CNAO) has been treating patients since 2011 with carbon-ion beams using the active-scanning modality. In such irradiation modality, the beam spot, which scans the treatment area, is characterised by very high particle-fluence rates (more than 105 s-1 mm-2). Moreover, the Bragg-peak is only ~1 mm-FWHM. Commercial tissue-equivalent proportional counters (TEPC), like the Far West Technologies LET-½, are large, hence they have limited capability to measure at high counting fluence rates. In this study we have used two home-made detectors, a mini-TEPC 0.81 mm2 in sensitive area and a silicon telescope 0.125 mm2 in sensitive area, to perform microdosimetric measurements in the therapeutic carbon-ion beam of CNAO. A monoenergetic carbon-ion beam of 189.5 ± 0.3 MeV/u scanning a 3 × 3 cm2 area has been used. Spectral differences are visible in the low y-value region, but the mean microdosimetric values, measured with the two detectors, result to be pretty consistent, as well as the microdosimetric spectra in the high y-value region.

MICRODOSIMETRIC STUDY AT THE CNAO ACTIVE-SCANNING CARBON-ION BEAM

Pola, A;BORTOT, DAVIDE;Fazzi, A;Agosteo, S;
2018-01-01

Abstract

The Italian National Centre for Oncological Hadrontherapy (CNAO) has been treating patients since 2011 with carbon-ion beams using the active-scanning modality. In such irradiation modality, the beam spot, which scans the treatment area, is characterised by very high particle-fluence rates (more than 105 s-1 mm-2). Moreover, the Bragg-peak is only ~1 mm-FWHM. Commercial tissue-equivalent proportional counters (TEPC), like the Far West Technologies LET-½, are large, hence they have limited capability to measure at high counting fluence rates. In this study we have used two home-made detectors, a mini-TEPC 0.81 mm2 in sensitive area and a silicon telescope 0.125 mm2 in sensitive area, to perform microdosimetric measurements in the therapeutic carbon-ion beam of CNAO. A monoenergetic carbon-ion beam of 189.5 ± 0.3 MeV/u scanning a 3 × 3 cm2 area has been used. Spectral differences are visible in the low y-value region, but the mean microdosimetric values, measured with the two detectors, result to be pretty consistent, as well as the microdosimetric spectra in the high y-value region.
2018
File in questo prodotto:
File Dimensione Formato  
A67.pdf

Accesso riservato

Descrizione: articolo
: Publisher’s version
Dimensione 574.26 kB
Formato Adobe PDF
574.26 kB Adobe PDF   Visualizza/Apri
11311-1047287_Bortot.pdf

accesso aperto

: Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione 730.14 kB
Formato Adobe PDF
730.14 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1047287
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 15
social impact