A method for the reconstruction of a vessel centerline from angiographic images is outlined in this work. A typical coronary artery segment with bifurcations was emulated with a 3D printed static phantom and several angiograms were acquired at various angular positions on the C-Arm. The effectiveness of the reconstruction turned out to be largely influenced by the intrinsic parameters of the angiographic system, particularly the homogeneous coordinates system scaling factor λ. Therefore, recourse was made to a heuristic optimization method to estimate the optimal value of λ for each view. We measured the reliability of the reconstruction method by varying the fitness function of the optimization step and measuring the distances of 8 test points in comparison to the corresponding points identified in the μCT centerline. Preliminary results showed that, with an adequate number of views, the adoption of the optimal fitness function allowed the median distance error to be decreased below the acceptance threshold of 10%. As expected, the reliability of the method is improved by increasing the number of processed views.

A method for coronary bifurcation centerline reconstruction from angiographic images based on focalization optimization

MONTIN, EROS;MIGLIORI, SUSANNA;CHIASTRA, CLAUDIO;CREDI, CATERINA;FEDELE, ROBERTO;LEVI, MARINELLA;MIGLIAVACCA, FRANCESCO;MAINARDI, LUCA
2016-01-01

Abstract

A method for the reconstruction of a vessel centerline from angiographic images is outlined in this work. A typical coronary artery segment with bifurcations was emulated with a 3D printed static phantom and several angiograms were acquired at various angular positions on the C-Arm. The effectiveness of the reconstruction turned out to be largely influenced by the intrinsic parameters of the angiographic system, particularly the homogeneous coordinates system scaling factor λ. Therefore, recourse was made to a heuristic optimization method to estimate the optimal value of λ for each view. We measured the reliability of the reconstruction method by varying the fitness function of the optimization step and measuring the distances of 8 test points in comparison to the corresponding points identified in the μCT centerline. Preliminary results showed that, with an adequate number of views, the adoption of the optimal fitness function allowed the median distance error to be decreased below the acceptance threshold of 10%. As expected, the reliability of the method is improved by increasing the number of processed views.
2016
Conf Proc IEEE Eng Med Biol Soc.
File in questo prodotto:
File Dimensione Formato  
Montin- IEEE-07591644.pdf

Accesso riservato

: Publisher’s version
Dimensione 905.57 kB
Formato Adobe PDF
905.57 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/1013247
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 1
social impact