Two categories of heart valve prostheses are currently available in clinics: mechanical and biological valves. The former show longer duration performance, on the other hand they induce high pressure drops during operation. The latter have preferable fluid dynamic performance but do not display high durability. The potential of Polymeric Heart Valves (PHVs) is to combine the haemodynamic properties of biological valves with the durability of mechanical valves. Our group has recently developed a new PHV made of a styrenic block copolymer obtained by compression moulding. Indeed, anisotropic mechanical properties of this material, mimicking the native tissue behaviour, can be obtained by compression and slow injection moulding. A computational tool was developed to optimize the PHV design and leaflet microstructure in order to reduce regions of stress concentration and increase the device lifetime. This work presents a hydrodynamic evaluation of the new PHV prototypes, which were manufactured by injection moulding based on the results obtained by the computational tool. The PHVs were tested under pulsatile flow conditions as prescribed by ISO 5840 Standard. Furthermore, continuous flow tests were performed to evaluate the velocity fields with the particle image velocimetry (PIV).

Fluid Dynamic assessment if new PHVs prototypes under pulsatile flow condition and using PIV

DE GAETANO, FRANCESCO;COSTANTINO, MARIA LAURA
2016-01-01

Abstract

Two categories of heart valve prostheses are currently available in clinics: mechanical and biological valves. The former show longer duration performance, on the other hand they induce high pressure drops during operation. The latter have preferable fluid dynamic performance but do not display high durability. The potential of Polymeric Heart Valves (PHVs) is to combine the haemodynamic properties of biological valves with the durability of mechanical valves. Our group has recently developed a new PHV made of a styrenic block copolymer obtained by compression moulding. Indeed, anisotropic mechanical properties of this material, mimicking the native tissue behaviour, can be obtained by compression and slow injection moulding. A computational tool was developed to optimize the PHV design and leaflet microstructure in order to reduce regions of stress concentration and increase the device lifetime. This work presents a hydrodynamic evaluation of the new PHV prototypes, which were manufactured by injection moulding based on the results obtained by the computational tool. The PHVs were tested under pulsatile flow conditions as prescribed by ISO 5840 Standard. Furthermore, continuous flow tests were performed to evaluate the velocity fields with the particle image velocimetry (PIV).
2016
Atti del V Congresso Gruppo Nazionale di Bioingegneria 2016
978-88-941906-0-1
Heart valve, styrenic block copolymer, pulse duplicator, particle image velocimetry.
File in questo prodotto:
File Dimensione Formato  
GNB_final_2.pdf

Accesso riservato

: Publisher’s version
Dimensione 910.97 kB
Formato Adobe PDF
910.97 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/995570
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact