Aims: This study sought to investigate the possible influence of different bifurcation stenting techniques on stent deformation, physical stress, and drug elution using a virtual tool that includes structural, fluid dynamics and drug-eluting numerical models. Methods and results: A virtual bench test based on explicit dynamics modelling was used to simulate procedures on bifurcated coronary vessels performed according to three different stenting techniques: provisional side branch stenting, culotte, and Tryton-based culotte. Geometrical configurations obtained after virtual stenting simulations were used to perform fluid dynamics and drug elution analyses. The results showed that substantially different patterns of mechanical deformation, shear stress and theoretical drug elution were obtained using the different techniques. Compared with conventional culotte, the dedicated Tryton seems to facilitate the intervention in terms of improved access to the main branch and to lower its biomechanical influence on the coronary bifurcation in terms of mechanical and haemodynamic parameters. However, since the Tryton stent is a bare metal stent, the drug elution obtained is lower. Conclusions: Numerical models might successfully complement the information on stenting procedures obtained with traditional approaches such as in vitro bench testing or clinical trials. Devices dedicated to bifurcations may facilitate procedure completion and may result in specific patterns of mechanical stress, regional blood flow and drug elution.

Stent deformation, physical stress, and drug elution obtained with provisional stenting, conventional culotte and Tryton-based culotte to treat bifurcations: a virtual simulation study

CHIASTRA, CLAUDIO;CUTRI', ELENA;ZUNINO, PAOLO;FORMAGGIA, LUCA;DUBINI, GABRIELE ANGELO;MIGLIAVACCA, FRANCESCO
2014-01-01

Abstract

Aims: This study sought to investigate the possible influence of different bifurcation stenting techniques on stent deformation, physical stress, and drug elution using a virtual tool that includes structural, fluid dynamics and drug-eluting numerical models. Methods and results: A virtual bench test based on explicit dynamics modelling was used to simulate procedures on bifurcated coronary vessels performed according to three different stenting techniques: provisional side branch stenting, culotte, and Tryton-based culotte. Geometrical configurations obtained after virtual stenting simulations were used to perform fluid dynamics and drug elution analyses. The results showed that substantially different patterns of mechanical deformation, shear stress and theoretical drug elution were obtained using the different techniques. Compared with conventional culotte, the dedicated Tryton seems to facilitate the intervention in terms of improved access to the main branch and to lower its biomechanical influence on the coronary bifurcation in terms of mechanical and haemodynamic parameters. However, since the Tryton stent is a bare metal stent, the drug elution obtained is lower. Conclusions: Numerical models might successfully complement the information on stenting procedures obtained with traditional approaches such as in vitro bench testing or clinical trials. Devices dedicated to bifurcations may facilitate procedure completion and may result in specific patterns of mechanical stress, regional blood flow and drug elution.
2014
File in questo prodotto:
File Dimensione Formato  
2014-Morlacchi-Eurointervention-71st_issue-article_242.pdf

Accesso riservato

: Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione 6.57 MB
Formato Adobe PDF
6.57 MB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/802518
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 25
  • ???jsp.display-item.citation.isi??? 21
social impact