A microfluidic thermal cycler for Polymerase Chain Reaction (PCR) has been modeled. A microliter sample is driven along the microchannel by a flow generated by means of electroosmotic micropumps, activated in sequence. Several multichannel pumps configurations have been separately modeled and the efficiency of each device has been evaluated with regard to the effective flowrate and the back flowrate. Results from these simulations have been then used to define the boundaries for the model of the entire device.

Model of a Microfluidic Thermal Cycler Activated by Means of Electro-Osmotic Micro-Pumps

BIANCHI, ELENA;CRITELLI, IDA;DUBINI, GABRIELE ANGELO
2012-01-01

Abstract

A microfluidic thermal cycler for Polymerase Chain Reaction (PCR) has been modeled. A microliter sample is driven along the microchannel by a flow generated by means of electroosmotic micropumps, activated in sequence. Several multichannel pumps configurations have been separately modeled and the efficiency of each device has been evaluated with regard to the effective flowrate and the back flowrate. Results from these simulations have been then used to define the boundaries for the model of the entire device.
2012
Comsol Conference 2012 Milan Proceedings (CD Rom)
9780983968870
File in questo prodotto:
File Dimensione Formato  
Comsol2012 EOF.pdf

Accesso riservato

: Post-Print (DRAFT o Author’s Accepted Manuscript-AAM)
Dimensione 404.26 kB
Formato Adobe PDF
404.26 kB Adobe PDF   Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/689952
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact