The two-hit resolution of a silicon drift chamber is measured using a pulsed Nd: YAG laser and a time digitizer readout. The data is analyzed by forming the covariance matrix in time samples, and transforming to a matrix in amplitude and time variation of each of the two hits. The resolution on the two-hit separation is found to be better than 25 μm with a drift field of 530 V/cm and a separation of more than 500 μm, with the resolution increasing to 50 μm as the separation nears 500 μ. Results are also presented for multiply ionizing tracks, showing a great improvement over single minimum ionizing

Double particle resolution measured in a silicon drift chamber

GATTI, EMILIO;SAMPIETRO, MARCO;
1992-01-01

Abstract

The two-hit resolution of a silicon drift chamber is measured using a pulsed Nd: YAG laser and a time digitizer readout. The data is analyzed by forming the covariance matrix in time samples, and transforming to a matrix in amplitude and time variation of each of the two hits. The resolution on the two-hit separation is found to be better than 25 μm with a drift field of 530 V/cm and a separation of more than 500 μm, with the resolution increasing to 50 μm as the separation nears 500 μ. Results are also presented for multiply ionizing tracks, showing a great improvement over single minimum ionizing
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11311/665876
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact