The evaluation of mechanical properties of aortic vascular prostheses was presented. A comparison among in vitro, theoretical and clinical data was done. The results showed that mean value and standard deviation of dilatation for the experimental measurements were 0.208±0.045 mm, while those of in vivo measurements were 0.209±0.046 mm.
Evaluation of Mechanical Properties of Aortic Vascular Prostheses: Comparison Among In Vitro, theoretical and Clinical Data.
VILLA, TOMASO MARIA TOBIA;MIGLIAVACCA, FRANCESCO;QUAGLINI, VIRGINIO;VENA, PASQUALE;CONTRO, ROBERTO;PIETRABISSA, RICCARDO
2001-01-01
Abstract
The evaluation of mechanical properties of aortic vascular prostheses was presented. A comparison among in vitro, theoretical and clinical data was done. The results showed that mean value and standard deviation of dilatation for the experimental measurements were 0.208±0.045 mm, while those of in vivo measurements were 0.209±0.046 mm.File in questo prodotto: